Abstract

The implementation of digital electronics for the multi-detector Gammasphere array has provided an opportunity to perform experiments which exceed the technical capabilities of its analog counterpart. The pulse shape analysis of HPGe detectors is presented for the purpose of determining the gamma-ray energy under high-counting rates and short integration times with the aim of improving the data throughput. A revised trapezoidal algorithm is delineated which is able to determine the gamma-ray energies during the offline analysis without being constrained by predetermined parameters. The performance of this algorithm is discussed, and compared with that of the analog system. The measured energy resolution meets expectations for operations with high-counting rates and short integration times.